Home FOR AUTHORS Bratislava Medical Journal 2017 Bratislava Medical Journal Vol.118, No.11, p.647-653, 2017

Journal info


 


Published Monthly, in English
Founded: 1919
ISSN 0006-9248
(E)ISSN 1336-0345

Impact factor 1.5

 

Select Journal







Webshop Cart

Your Cart is currently empty.

Info: Your browser does not accept cookies. To put products into your cart and purchase them you need to enable cookies.

Bratislava Medical Journal Vol.118, No.11, p.647-653, 2017

Title: Prevalent placement error of deep brain stimulation electrode in movement disorders (technical considerations)
Author: M. Kloc, Z. Kosutzka, J. Steno, P. Valkovic

Abstract: BACKGROUND: Deep brain stimulation is an effective and safe technique. Displacement of the electrode relative to the optimal stimulation site can lead to insufficient effect and sometimes to the need of operative electrode re-position.
OBJECTIVE: This study was aimed to analyse targeting accuracy of deep brain stimulation electrode implantation to subthalamic nucleus (STN) and globus pallidus internus (Gpi). It detected possible causes of inaccuracy and prevalent shift to certain direction.
METHODS: Targeting accuracy was analysed in 47 patients with Parkinson´s disease (PD) and 11 patients with dystonia with bilateral implantation of deep brain stimulation electrodes between years 2009 and 2016.
RESULTS: A shift of electrode to prevalent direction was observed on the left side to medial and posterior and on the right side to lateral direction. Greater shift was observed on the left side and in a higher angulation of trajectory laterally. Movement of the electrode, because of its traction in anchoring device, was identified as a possible factor for prevalent electrode shift. Calibration of stereotactic coordinates to correct prevalent shift was used.
CONCLUSION: Targeting inaccuracy is the result of accumulation of errors in individual steps of electrode implantation. Direction of the shift can be random or it can be toward a prevalent direction. A correction of prevalent error can prevent a suboptimal electrode placement (Tab. 3, Fig. 11, Ref. 29).

Keywords: deep brain stimulation, calibration, targeting accuracy, stereotactic techniques
Published online: 05-Dec-2017
Year: 2017, Volume: 118, Issue: 11 Page From: 647, Page To: 653
doi:10.4149/BLL_2017_123


download file



© AEPress s.r.o
Copyright notice: For any permission to reproduce, archive or otherwise use the documents in the ELiS, please contact AEP.